Header
home  |  about us  |  unique support  |  users  |  prices  |  newsletters  |  contact us  |  Adv. Physics Lab Assoc.  |  events
  Navigation


Pulsed NMR

New Pulsed/CW NMR Designed for Teaching

Pulsed NMR PS1-B Specifications

Magnet
  • Field: 3,600 Gauss (nominal)
  • Homogeneity: 0.1 % over 1 cm³ spherical volume
  • Temperature stability: 4 Gauss/°C

Pulse Programmer
  • Two Independent Pulses: A and B
  • Pulse width: 1-30 µs
  • Pulse sequence: Single A, Multiple B; 1-99 pulses
  • Digital time delay: 0.1 ms-9.9 s in 10 µs increments
  • Repetition time: 1 ms-10 s
  • Internal, external, manual trigger
  • Meiboom-Gill gate pulse
  • Scope trigger on A or B pulse

Oscillator/Amplifier/Mixer
  • Frequency: 15.000 ± 1.0 MHz, 10 Hz resolution
  • Accuracy: 0.005 %
  • Stability: 1 ppm over 103 s
  • Leakage (pulse off): 80 db maximum
  • B rotating 10 Gauss, 5 µs for p/2 pulse
  • Mixer input 10 mV rms full-scale

Receiver
  • Sensitivity: 8 µV rms full-scale
  • Equivalent input noise: 1.5 µV rms
  • Gain range: 60 db (power)
  • Dead time: 5 µs
  • Time constant: 0.01-0.3 ms
  • RF output, detected output, blanking input

Sample Probe
  • Crossed-coil design: Helmholtz configuration for transmitter, axial receiver coil
  • Probe mounted on mechanical stage to locate sample in appropriate region of the magnet gap
  • Two-dimensional (x-y) motion, parallel to pole pieces.

Instrument Case
  • Power supply, two open slots for additional accessories.

Also included
  • Complete set of coaxial cables,
  • 50 sample vials, vial storage case and standard calibrated samples.
  • Instructional manual with representive data, student experiments and extended references.
Accessories Available
  • Circuit Diagrams
  • B1 - Pickup Probe to measure rf magnetic field diretly
  • Dummy Sample Probe to simulate precessing nuclear magnetism
  • Carry Handle for Electronics Case