Earth's Field Nuclear Magnetic Resonance (NMR) Gradient/Field Coil System
Newsletter 1 – Transforming Earth's Field NMR
Newsletter 2 – Putting the 'R' Into the Earth's Field NMR
Learn about Earth's Field NMR
Gradient/Field Coil System Brochure
Earth's Field NMR Brochure
Specifications
Helmholtz Coils
30 Turns/coil, #20 AWG Copper Wire
Average Coil Radius: 30.23 cm
Total Series Resistance: 3.8W
Maximum Continuous Current: 3A
Coil Constant: 89 µT/A
Uniformity: 0.01% over sample volume
Gradient Coils
Range: ±5 µT/m
Calibration Constant: 250 µT/m per Amp
Controller
Maximum gradient current: ±20 mA
Step delay time: 0-2.5 s
Maximum step cur rent: ±10 mA
Learn about Earth's Field NMR.
introduction | the
instrument | experiments | specifications
| accessories
| prices
|
 |