![]() |
|
|
Diode Laser Spectroscopy
CHARACTERISTICS OF THE TUNABLE DIODE LASER With the grating feedback in place, students can observe the laser's wavelength stability, the frequency sweep (using both grating angle and current modulation), and the sweep interruptions due to mode hopping. It is also possible to change the external cavity length and measure its effect. The simplest form of optical spectroscopy, is shown below in Figure 1.
The frequency swept laser light is passed through a cell containing rubidium vapor and the transmitted light is detected. Data for natural rubidium is shown in the oscilloscope capture below. The broad absorption peaks observed by this method correspond to transitions between the ground S1/2 and the exited P3/2 states of both isotopes of rubidium.
SATURATED ABSORPTION SPECTROSCOPY
Given the DF = 0, ±1 selection rules, the energy level diagram of 85Rb in Figure 3 suggests that there should be on three, not six, transitions from F = 2. The six peaks seen in Figure 6 include an additional three crossover transition peaks. These additional peaks, at frequencies exactly halfway between pairs of "actual" transition frequencies, arise from atoms moving at non-zero velocities such that the pump is in resonance with one transition and the probe is in resonance with the other transition. USING THE INTERFEROMETER TO CALIBRATE THE SWEEP In order to make quantitative measurements of the hyperfine splittings and compare these measurements to the handbook date, it is essential to calibrate the frequency sweep of the laser. This is accomplished with an unequal-arm interferometer shown the the upper section of the block diagram of Figure 4. Beam splitter 2 diverts a small portion of the laser light into the interferometer assembly. The interferometer beam splitter divides the light, sending it to mirrors 1 and 2 along the long and short arms of the Michelson interferometer. Returning beams recombine and interfere at the beam splitter. Because of the unequal arm lengths, the frequency sweep of the laser generates a series of fringes in time, due to alternate constructive and destructive interference. Figure 7 shows a typical interference signal seen by Detector 3 along with the Doppler broadened transmission data.
If DL = L1- L2 is the is the difference between the one-way lengths of the two arms, then the optical frequency difference, df, between two successive maxima at the interferometer can be calculated as: df = c/2 DL. For our optical arrangement, with a path length difference of 0.35 meters, the sweep calibration is 0.429 GHz/fringe. This gives a 6.65 GHz frequency difference between the cursor marked features of Figure 7. The accepted value, shown on the energy diagram in Figure 3, is 6.835 Hz. But this is by no means the end of the story. This unit can measure Zeeman splittings of the excited states, Faraday rotation in rubidium vapor, and the refractive index of rubidium as well as the Clausius-Claperyon relationship in rubidium.
|